Owner’s ManualForm No. 702718 Rev. A 03/05Regal Intensive Care UnitControlled EnvironmentTemperature Controlled Air FlowFiltered, Re-circul
viiiSSCI - Suburban Surgical Company, Inc.Table of ContentsChapter 6 - Troubleshooting ...137General ...
88 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsOxygen SensorP/N 853395The oxygen sensor is connected, via a spiral cor
Form No. 702718 Rev. A March, 2005 89Owner’s Manual - Regal Intensive Care Unit4. Remove the fuel cell.5. Disposition of the fuel cell: If yo
90 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsInstallation1. Connect the white and black wires fromthe spiral cord of
Form No. 702718 Rev. A March, 2005 91Owner’s Manual - Regal Intensive Care Unit4. Unplug the spiral cord from the oxygenconcentration meter (F
92 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsOxygenConcentration MeterP/N 853436The oxygen concentration meter is mo
Form No. 702718 Rev. A March, 2005 93Owner’s Manual - Regal Intensive Care UnitNote: Terminal numbers are displayed on the rear of the meter
94 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsInstallation1. Put the new meter into its mounting hole in the door so
Form No. 702718 Rev. A March, 2005 95Owner’s Manual - Regal Intensive Care UnitKnobP/N 853461Three knobs are used on the electronics compartme
96 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsPotentiometer,CalibrationP/N 853397The potentiometer operates the calib
Form No. 702718 Rev. A March, 2005 97Owner’s Manual - Regal Intensive Care Unit7. Pull the knob off the shaft.8. With a 1/2-in. wrench, remov
Form No. 702718 Rev. A March, 2005 ixOwner’s Manual - Regal Intensive Care UnitSECTION 11 - TOXICOLOGICAL INFORMATION ...
98 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsCircuit BreakerP/N 853466Two identical 10 amp. circuit breakers are loc
Form No. 702718 Rev. A March, 2005 99Owner’s Manual - Regal Intensive Care Unit3. Replace the wire spade connectors on the circuit breakerterm
100 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & Replacements5. Pull all the wire spade connectors off the rear of the switch(Figur
Form No. 702718 Rev. A March, 2005 101Owner’s Manual - Regal Intensive Care UnitOn/Off Button(Floor Temperature,Left & Right)P/N 212242The
102 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & Replacements5. Un-plug the green wire from Terminal 6 on the rear of thetemperatur
Form No. 702718 Rev. A March, 2005 103Owner’s Manual - Regal Intensive Care Unit3. Pass the connectors and wires through the knurled ring ands
104 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsSystem LightP/N 854589There are six red system lights on the electroni
Form No. 702718 Rev. A March, 2005 105Owner’s Manual - Regal Intensive Care UnitDigital Display(Floor Temperature,Left & Right)P/N 853843T
106 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & Replacements6. Remove the printed circuit board/display assembly from therear of t
Form No. 702718 Rev. A March, 2005 107Owner’s Manual - Regal Intensive Care UnitFuse(Floor Temperature,Left & Right)P/N 853848Two fuses ar
xSSCI - Suburban Surgical Company, Inc.Table of ContentsNotes:
108 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsFuse Holder,(Floor Temperature,Left & Right)P/N 853671There are tw
Form No. 702718 Rev. A March, 2005 109Owner’s Manual - Regal Intensive Care UnitInstallation1. Insert the fuse holder through the door from th
110 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & Replacements* Wire colors can varyTable 4. Wire Terminations - Floor Temperature
Form No. 702718 Rev. A March, 2005 111Owner’s Manual - Regal Intensive Care Unit14. With a 1/2-in. wrench, remove the nut behind the knob.15.
112 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsPanel LatchP/N 853561There are four latches on the electronics compart
Form No. 702718 Rev. A March, 2005 113Owner’s Manual - Regal Intensive Care UnitThermostatP/N 853421CAUTION: Before proceeding, make sure the
114 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & Replacements9. Remove the plastic cover from thethermostat.10. Using tags, masking
Form No. 702718 Rev. A March, 2005 115Owner’s Manual - Regal Intensive Care UnitFigure 129. Capillary Tubing and BulbHolderWire Tie MountCapi
116 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsAudible AlarmP/N 85345811. Mount the thermostat and bracket to thepart
Form No. 702718 Rev. A March, 2005 117Owner’s Manual - Regal Intensive Care Unit7. Rotate the knurled ring on the audible alarm counter-clockw
Form No. 702718 Rev. A March, 20051Owner’s Manual - Regal Intensive Care UnitChapter 1 - General InformationIntroductionSSCI’s Regal Intensiv
118 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsSolenoid ValveP/N 8534401. Make sure that the main on/off switch isOff
Form No. 702718 Rev. A March, 2005 119Owner’s Manual - Regal Intensive Care UnitFigure 136. Solenoid Valve2 Wires toMain Terminal BlockOxygen
120 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsCAUTION: This procedure requires opening theoxygen system, and should
Form No. 702718 Rev. A March, 2005 121Owner’s Manual - Regal Intensive Care UnitFigure 138. Oxygen Flow Meter and Related PiecesBracketSuppor
122 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsRe-assembly1. Hold the bracket in place on the flow meter, re-attach t
Form No. 702718 Rev. A March, 2005 123Owner’s Manual - Regal Intensive Care UnitFan, Ventilating(ElectronicsCompartment)P/N 853414To replace o
124 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & Replacements3. Mount the fan in place and secure with the four screws.4. Mount the
Form No. 702718 Rev. A March, 2005 125Owner’s Manual - Regal Intensive Care UnitOctal Base(Alarm TimeDelay Relay)P/N 853871CAUTION: Before pr
126 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsRelayP/N 853415CAUTION: Before proceeding, make sure the unit isOff,
Form No. 702718 Rev. A March, 2005 127Owner’s Manual - Regal Intensive Care UnitDigital ReadoutP/N 854591Tool Required Phillips screwdriverRe
2SSCI - Suburban Surgical Company, Inc.Chapter 1 - General InformationWARNINGSThroughout this manual you will find text under the headings Note,CAUTIO
128 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsPanel, Filter/Fans,CompleteP/N 206874Procedure1. Turn all four panel a
Form No. 702718 Rev. A March, 2005 129Owner’s Manual - Regal Intensive Care UnitGasket, Filter/FansCompartmentP/N 754456There are two sets of
130 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsFilter, Filter/FansCompartmentP/N 853469Procedure1. Open the filter/fa
Form No. 702718 Rev. A March, 2005 131Owner’s Manual - Regal Intensive Care UnitSodalime PanP/N 206537CAUTION: Sodalime can cause burns to ey
132 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsHeater, ElectricP/N 853455CAUTION: Before proceeding, make sure the u
Form No. 702718 Rev. A March, 2005 133Owner’s Manual - Regal Intensive Care Unit9. The rear of the heater is mounted to a bracket similar toth
134 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsFan, Circulating(Filter/FansCompartment)P/N 853414To replace the venti
Form No. 702718 Rev. A March, 2005 135Owner’s Manual - Regal Intensive Care Unit10. With a Phillips screwdriver, remove the four screws thatho
136 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsNotes:
Form No. 702718 Rev. A March, 2005137Owner’s Manual - Regal Intensive Care UnitChapter 6 - TroubleshootingGeneralThe following procedures wil
Form No. 702718 Rev. A March, 20053Owner’s Manual - Regal Intensive Care UnitThe following accessories increase the operating convenience ofy
138 SSCI - Suburban Surgical Company, Inc.Chapter 6 - TroubleshootingReturning theIntensive Care Unitfor RepairsRMA NumbersPacking andShipmentPage num
Form No. 702718 Rev. A March, 2005139Owner’s Manual - Regal Intensive Care UnitThe ICU will not turn on.Note: When you press the main on/off
140 SSCI - Suburban Surgical Company, Inc.Chapter 6 - TroubleshootingCircuit breakers keep tripping.First: If this is a new installation, does it mee
Form No. 702718 Rev. A March, 2005141Owner’s Manual - Regal Intensive Care UnitOxygen flow is inadequate.First: Check your oxygen supply. Is
142 SSCI - Suburban Surgical Company, Inc.Chapter 6 - TroubleshootingOxygen flow is excessive.First: Make sure your oxygen pressure regulator is corr
Form No. 702718 Rev. A March, 2005143Owner’s Manual - Regal Intensive Care UnitAlarm keeps going off.First: Make sure that all four panel ac
144 SSCI - Suburban Surgical Company, Inc.Chapter 6 - TroubleshootingThe alarm system does not work.First: Make sure that the main on/off switch and
Form No. 702718 Rev. A March, 2005145Owner’s Manual - Regal Intensive Care UnitHeat will not come on or will not hold constant.First: Make s
146 SSCI - Suburban Surgical Company, Inc.Chapter 6 - TroubleshootingCooling will not come on or will not hold constant.First: Make sure the temperat
Form No. 702718 Rev. A March, 2005147Owner’s Manual - Regal Intensive Care UnitFloor heat doesn’t work or will not hold constant.There are tw
4SSCI - Suburban Surgical Company, Inc.Chapter 1 - General InformationCAUTION: Before working on the electrical system,make sure the unit is Off and
148 SSCI - Suburban Surgical Company, Inc.Chapter 6 - TroubleshootingVentilating fan does not work.First: Make sure that electrical power to the unit
Form No. 702718 Rev. A March, 2005149Owner’s Manual - Regal Intensive Care UnitCirculating fan does not work.First: Make sure that electrica
150 SSCI - Suburban Surgical Company, Inc.Chapter 6 - TroubleshootingNo readings on digital readout.First: The battery is probably dead. Refer to Pa
Form No. 702718 Rev. A March, 2005151Owner’s Manual - Regal Intensive Care UnitExcessive carbon dioxide in the animal compartment.First: Th
152 SSCI - Suburban Surgical Company, Inc.Chapter 6 - TroubleshootingNo reading on oxygen concentration meter.First: You may have an exhausted fuel c
Form No. 702718 Rev. A March, 2005153Owner’s Manual - Regal Intensive Care UnitCalibration knob will not move the calibration needleThis is a
154 SSCI - Suburban Surgical Company, Inc.Chapter 6 - TroubleshootingCannot move the set-point needle in the oxygenconcentration meter.CAUTION: Befor
Form No. 702718 Rev. A March, 2005155Owner’s Manual - Regal Intensive Care UnitFloor temperature fuse keeps blowing.CAUTION: Before working
156 SSCI - Suburban Surgical Company, Inc.Chapter 6 - TroubleshootingNotes:
Form No. 702718 Rev. A March, 2005A-1Owner’s Manual - Regal Intensive Care UnitAppendixThis Appendix contains Material Data Safety Sheets (MS
Form No. 702718 Rev. A March, 20055Owner’s Manual - Regal Intensive Care UnitCare and Cleaning ofStainless SteelIntroductionCleaning andClean
A-2 SSCI - Suburban Surgical Company, Inc.AppendixThis page left blank intentionally.
Form No. 702718 Rev. A March, 2005A-3Owner’s Manual - Regal Intensive Care UnitPuritan Medical ProductsPuritan Medical Products, Inc.9101 Bon
A-4 SSCI - Suburban Surgical Company, Inc.AppendixSECTION 5 - FIRE FIGHTING MEASURESFlash Point: N/AFlammable Limit: N/AAutoignition: N/AExtinguishing
Form No. 702718 Rev. A March, 2005A-5Owner’s Manual - Regal Intensive Care UnitSECTION 12 - ECOLOGICAL INFORMATIONNo adverse ecological effec
A-6 SSCI - Suburban Surgical Company, Inc.AppendixThis page left blank intentionally.
Form No. 702718 Rev. A March, 2005A-7Owner’s Manual - Regal Intensive Care UnitHoneywellMaterial Safety Data SheetGenetron® MP39 (R-401A)1.
HoneywellMATERIAL SAFETY DATA SHEETGenetron® MP39 (R-401A)MSDS Number: GTRN-0006Current Issue Date: January, 2000A-8 SSCI - Suburban Surgical Company,
HoneywellMATERIAL SAFETY DATA SHEETGenetron® MP39 (R-401A)MSDS Number: GTRN-0006Current Issue Date: January, 2000Form No. 702718 January, 2005A-9Ow
HoneywellMATERIAL SAFETY DATA SHEETGenetron® MP39 (R-401A)MSDS Number: GTRN-0006Current Issue Date: January, 2000A-10 SSCI - Suburban Surgical Company
HoneywellMATERIAL SAFETY DATA SHEETGenetron® MP39 (R-401A)MSDS Number: GTRN-0006Current Issue Date: January, 2000Form No. 702718 January, 2005A-11O
6SSCI - Suburban Surgical Company, Inc.Chapter 1 - General InformationCleaningRequirementsEffect on WarrantyThe large selection of brands and combinat
HoneywellMATERIAL SAFETY DATA SHEETGenetron® MP39 (R-401A)MSDS Number: GTRN-0006Current Issue Date: January, 2000A-12 SSCI - Suburban Surgical Company
HoneywellMATERIAL SAFETY DATA SHEETGenetron® MP39 (R-401A)MSDS Number: GTRN-0006Current Issue Date: January, 2000Form No. 702718 January, 2005A-13O
HoneywellMATERIAL SAFETY DATA SHEETGenetron® MP39 (R-401A)MSDS Number: GTRN-0006Current Issue Date: January, 2000A-14 SSCI - Suburban Surgical Company
Inside back coverThis page left blank intentionally.
Form No. 702718 Rev. A 03/05Suburban Surgical Company, Inc.275 Twelfth Street Wheeling, IL 60
Form No. 702718 Rev. A March, 20057Owner’s Manual - Regal Intensive Care UnitSSCI ContactInformationSSCI Customer Service can be contacted vi
© Copyright 2005 by Suburban Surgical Co., Inc. All rights reserved.No part of this document may be reproduced or utilized in any form or by any means
8SSCI - Suburban Surgical Company, Inc.Chapter 1 - General InformationNotes:
Form No. 702718 Rev. A March, 20059Owner’s Manual - Regal Intensive Care UnitChapter 2 - Description of ComponentsOverviewThe sections in this
10 SSCI - Suburban Surgical Company, Inc.Chapter 2 - Description of ComponentsIntensive Care Unit,Front(Figure 3)1 - IV Access PortsFour IV access por
Form No. 702718 Rev. A March, 200511Owner’s Manual - Regal Intensive Care Unit3 - Door LatchesEach Intensive Care Unit door is held closed by
12 SSCI - Suburban Surgical Company, Inc.Chapter 2 - Description of Components8 - Air Outlet Grill7 - RemovablePVC Floor5 - DoorsThe full-size compart
Form No. 702718 Rev. A March, 200513Owner’s Manual - Regal Intensive Care Unit9 - OxygenConcentration MeterElectronicsCompartment,External(Fig
14 SSCI - Suburban Surgical Company, Inc.Chapter 2 - Description of Components10 - Calibration Knob11 - Set-pointAdjustment KnobTurning the set-point
Form No. 702718 Rev. A March, 200515Owner’s Manual - Regal Intensive Care UnitThe oxygen on/off switch (Figures 11 and 14) controls the flow o
16 SSCI - Suburban Surgical Company, Inc.Chapter 2 - Description of Components16 - Cooling On/OffRocker Switch &System Light17, 18 - Left & Ri
Form No. 702718 Rev. A March, 200517Owner’s Manual - Regal Intensive Care UnitOn/Off ButtonAn amber light in the switch illuminates when the s
Form No. 702718 Rev. A March, 2005 iOwner’s Manual - Regal Intensive Care UnitTable of ContentsChapter 1 - General Information ...
18 SSCI - Suburban Surgical Company, Inc.Chapter 2 - Description of Components21 - Audible AlarmTo access the interior of the electronics compartment,
Form No. 702718 Rev. A March, 200519Owner’s Manual - Regal Intensive Care Unit23 - OxygenFlow Meter andSolenoid Valve24 - Circuit BreakersThe
20 SSCI - Suburban Surgical Company, Inc.Chapter 2 - Description of ComponentsElectronicsCompartment,Internal(Figure 23)CAUTION: Before working on th
Form No. 702718 Rev. A March, 200521Owner’s Manual - Regal Intensive Care UnitA ventilating fan (Figures 23 and 24) provides a steady flow of
22 SSCI - Suburban Surgical Company, Inc.Chapter 2 - Description of ComponentsThe time delay relay (Figures 23 and 25) determines how soon thealarm wi
Form No. 702718 Rev. A March, 200523Owner’s Manual - Regal Intensive Care Unit34 - Oxygen Port toFilter/FansCompartmentAfter leaving the oxyge
24 SSCI - Suburban Surgical Company, Inc.Chapter 2 - Description of ComponentsFilter/FansCompartment,External35 - Digital ReadoutThis two-function dig
Form No. 702718 Rev. A March, 200525Owner’s Manual - Regal Intensive Care Unit37 - Panel AccessKnobsRotate the four panel access knobs one-qua
26 SSCI - Suburban Surgical Company, Inc.Chapter 2 - Description of ComponentsFilter/FansCompartment,Internal(Figure 29)The following section gives yo
Form No. 702718 Rev. A March, 200527Owner’s Manual - Regal Intensive Care UnitAfter leaving the oxygen flow meter, oxygen is directed throughp
iiSSCI - Suburban Surgical Company, Inc.Table of Contents13 - Oxygen On/Off Rocker Switch & System Light ...
28 SSCI - Suburban Surgical Company, Inc.Chapter 2 - Description of ComponentsThe chiller (Figures 29 and 31) cools the circulating air as it passesth
Form No. 702718 Rev. A March, 200529Owner’s Manual - Regal Intensive Care Unit43 - Circulating FansThree circulating fans (Figures 29 and 33)
30 SSCI - Suburban Surgical Company, Inc.Chapter 2 - Description of ComponentsIntensive Care Unit,Rear(Figure 34)The following section gives you detai
Form No. 702718 Rev. A March, 200531Owner’s Manual - Regal Intensive Care UnitA ventilating fan provides a steady flow of air through the elec
32 SSCI - Suburban Surgical Company, Inc.Chapter 2 - Description of ComponentsNotes:
Form No. 702718 Rev. A March, 2005 33Owner’s Manual - Regal Intensive Care UnitChapter 3 - Installation & SetupUnpacking andInspectionCAUT
34 SSCI - Suburban Surgical Company, Inc.Chapter 3 - Installation & SetupInstalling and setting up the Regal Intensive Care Unit is a twelvestep p
Form No. 702718 Rev. A March, 2005 35Owner’s Manual - Regal Intensive Care UnitInstalling theFuel CellA fuel cell in the oxygen sensor powers
36 SSCI - Suburban Surgical Company, Inc.Chapter 3 - Installation & SetupFigure 39. Oxygen Concentration Meterand Calibration KnobBlack Calibrati
Form No. 702718 Rev. A March, 2005 37Owner’s Manual - Regal Intensive Care Unit5. Gently, screw the sensor end cap (Figure 38) back on. Becare
Form No. 702718 Rev. A March, 2005 iiiOwner’s Manual - Regal Intensive Care UnitChapter 3 - Installation & Setup ...
38 SSCI - Suburban Surgical Company, Inc.Chapter 3 - Installation & SetupConnectingElectrical PowerCAUTION: The Regal Intensive Care Unit require
Form No. 702718 Rev. A March, 2005 39Owner’s Manual - Regal Intensive Care UnitChecking theHeating FunctionNote: To speed up these tests, ins
40 SSCI - Suburban Surgical Company, Inc.Chapter 3 - Installation & Setup1. Push the main on/off switch (Figure 40) to On. The amberlight in the s
Form No. 702718 Rev. A March, 2005 41Owner’s Manual - Regal Intensive Care UnitChecking FloorHeat (Left)1. Push the main on/off switch (Figure
42 SSCI - Suburban Surgical Company, Inc.Chapter 3 - Installation & SetupChecking theOxygen SystemNote: You can perform this test even though oxy
Form No. 702718 Rev. A March, 2005 43Owner’s Manual - Regal Intensive Care UnitThe alarm system on the Regal Intensive Care Unit warns you ift
44 SSCI - Suburban Surgical Company, Inc.Chapter 3 - Installation & Setup10. Push the alarm on/off switch to Off. The light in the switchgoes out.
Form No. 702718 Rev. A March, 2005 45Owner’s Manual - Regal Intensive Care UnitFilling theCO2 Filter6. Push the oxygen on/off switch to On. Th
46 SSCI - Suburban Surgical Company, Inc.Chapter 3 - Installation & SetupDisposition of theShipping Carton To fill the Sodalime Pan:1. Open the fi
Form No. 702718 Rev. A March, 200547Owner’s Manual - Regal Intensive Care UnitChapter 4 - Operation & CareOperating the SSCI Regal Intensi
ivSSCI - Suburban Surgical Company, Inc.Table of ContentsControlling Oxygen Pressure ...
48SSCI - Suburban Surgical Company, Inc.Chapter 4 - Operation and CareOpening and Closingthe DoorsThe Intensive Care Unit has two doors that are each
Form No. 702718 Rev. A March, 200549Owner’s Manual - Regal Intensive Care UnitUsing the DividerPanelsGeneralTwo removable divider panels are s
50SSCI - Suburban Surgical Company, Inc.Chapter 4 - Operation and Care3. Orient the divider panel so that the notched corner is ontop and toward the f
Form No. 702718 Rev. A March, 200551Owner’s Manual - Regal Intensive Care UnitUsing OxygenWARNING: Oxygen is a highly combustible gas.Avoid t
52SSCI - Suburban Surgical Company, Inc.Chapter 4 - Operation and CareTo stop the oxygen flow, press the oxygen on/off switch. The lightin the switch
Form No. 702718 Rev. A March, 200553Owner’s Manual - Regal Intensive Care UnitHeating theCompartment AirSince the temperature in the room in w
54SSCI - Suburban Surgical Company, Inc.Chapter 4 - Operation and CareIf cooling is required by animals in both half-compartments, usethe wire grill d
Form No. 702718 Rev. A March, 200555Owner’s Manual - Regal Intensive Care UnitUsing the AlarmSystemThe alarm system alerts you, with audible a
56SSCI - Suburban Surgical Company, Inc.Chapter 4 - Operation and CareControlling OxygenPressureControlling OxygenFlow RateFigure 59. Oxygen Flow Met
Form No. 702718 Rev. A March, 200557Owner’s Manual - Regal Intensive Care UnitAccessing theElectronicsCompartmentFigure 60. Electronics Compa
Form No. 702718 Rev. A March, 2005 vOwner’s Manual - Regal Intensive Care UnitLatch Connector Holder ...
58SSCI - Suburban Surgical Company, Inc.Chapter 4 - Operation and CareOpening the filter/fans compartment panel exposeselectrical components inside. U
Form No. 702718 Rev. A March, 200559Owner’s Manual - Regal Intensive Care UnitSetting the AlarmTime Delay RelayThis component determines how s
60SSCI - Suburban Surgical Company, Inc.Chapter 4 - Operation and CareRe-setting theCircuit BreakersTwo 10 amp. circuit breakers are provided on the f
Form No. 702718 Rev. A March, 200561Owner’s Manual - Regal Intensive Care UnitFigure 70. Fuse Cap and FuseFuse CapFuseFigure 69. Floor Temper
62SSCI - Suburban Surgical Company, Inc.Chapter 4 - Operation and CareReplacing the DigitalReadout BatteryThe digital readout operates independently o
Form No. 702718 Rev. A March, 200563Owner’s Manual - Regal Intensive Care UnitCleaning the RegalIntensive Care UnitOrdinary deposits of waste
64SSCI - Suburban Surgical Company, Inc.Chapter 4 - Operation and CarePreventiveMaintenance for YourRegal Intensive CareUnitIntroductionThe Regal Inte
Form No. 702718 Rev. A March, 200565Owner’s Manual - Regal Intensive Care UnitCAUTION: Sodalime can cause burns to eyes andskin. Dust can cau
66SSCI - Suburban Surgical Company, Inc.Chapter 4 - Operation and CareSodalimePanFigure 73. Sodalime Pan in Filter/FansCompartment4. Open a single co
Form No. 702718 Rev. A March, 200567Owner’s Manual - Regal Intensive Care UnitCleaning the Filter/Fans CompartmentFilterA metal mesh filter in
viSSCI - Suburban Surgical Company, Inc.Table of ContentsCircuit Breaker ...
68SSCI - Suburban Surgical Company, Inc.Chapter 4 - Operation and CareReplacing theFuel CellCAUTION: Do not remove the fuel cell from the baguntil re
Form No. 702718 Rev. A March, 200569Owner’s Manual - Regal Intensive Care UnitFigure 81. Oxygen Concentration Meterand Calibration KnobBlack
70SSCI - Suburban Surgical Company, Inc.Chapter 4 - Operation and CareCleaning theFilter/Fans andElectronicsCompartmentsNew Style Units1. Do not remov
Form No. 702718 Rev. A March, 2005 71Owner’s Manual - Regal Intensive Care UnitChapter 5 - Repairs & ReplacementsReplacement Parts Table 3
72 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsTable 3. Replacement Parts for the Regal Intensive Care UnitemaNtraP r
Form No. 702718 Rev. A March, 2005 73Owner’s Manual - Regal Intensive Care Unit Many of the threaded fasteners used on SSCI products aresecur
74 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsParts ReplacementProceduresCompartment DoorLeft - P/N 205994Right - P/N
Form No. 702718 Rev. A March, 2005 75Owner’s Manual - Regal Intensive Care UnitDoor HingeLeft Door - P/N 212035Right Door - P/N 212036There is
76 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsDoor Gasket P/N 754455The gasket on each animal compartment door consis
Form No. 702718 Rev. A March, 2005 77Owner’s Manual - Regal Intensive Care UnitLatch ConnectorRest P/N 618472There is a latch connector rest o
Form No. 702718 Rev. A March, 2005 viiOwner’s Manual - Regal Intensive Care UnitOxygen Flow Meter ...
78 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsLatch ConnectorHolderP/N 618479The latch connector holders hold the lat
Form No. 702718 Rev. A March, 2005 79Owner’s Manual - Regal Intensive Care UnitLatch ConnectorP/N 212047There is one latch connector on each a
80 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsLatch BracketTop - P/N 618480Bottom - P/N 618481There is one top and on
Form No. 702718 Rev. A March, 2005 81Owner’s Manual - Regal Intensive Care UnitDivider PanelWire Grill - P/N 206001Clear Acrylic - P/N 749575T
82 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsSeal, IV PortP/N 752100There are four IV ports on the Intensive Care Un
Form No. 702718 Rev. A March, 2005 83Owner’s Manual - Regal Intensive Care UnitElectric Power CordP/N 8534312. Unplug the power cord from the
84 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsRemovable PCV FloorP/N 102275-00Two orange, removable PCV floors are pr
Form No. 702718 Rev. A March, 2005 85Owner’s Manual - Regal Intensive Care UnitFuel CellP/N 853396Figure 95. Oxygen Sensor Inside theElectron
86 SSCI - Suburban Surgical Company, Inc.Chapter 5 - Repairs & ReplacementsFigure 98. Oxygen Concentration Meterand Calibration KnobBlack Calibra
Form No. 702718 Rev. A March, 2005 87Owner’s Manual - Regal Intensive Care UnitNew Style Units1. Do not remove the oxygen sensor from the part
Comentários a estes Manuais